R&S ZVL GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
233
Transformation networks can be defined by a set of S-parameters stored in a Touchstone file or
b
y an equivalent circuit with lumped elements.
The same networks are available for embedding and deembedding.
De-embedding opens the configuration dialog and activates or deactivates deembedding.
Embedding opens the configuration dialog and activates or deactivates embedding.
Embedding/Deembedding
Selects a 2-port transformation network for (de-)embedding, defines its parameters, assigns it to a
physical port and enables (de-)embedding. The two dialogs for deembedding and embedding are identical
except for their inverse effect.
The dialogs contain the following control elements:
Port is the analyzer port for the added or removed circuit. The transformation networks are
defined such that the analyzer test port is connected to the left of the circuit; the DUT port is on
the right side.
(De-)embed DUT enables or disables the (de-)embedding function.