R&S ZVL GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
241
The Calibration Kits dialog contains a series of buttons, each of them opening a dialog:
• Import Kit imports a cal kit file containing the parameters for a new calibration kit. It is possible to
load several kits, so the analyzer asks whether to set the imported calibration kit active.
• Add Kit combines different standards to form a new calibration kit.
• View/ Modify Kit adds or deletes standards in an imported or user-defined calibration kit and/or
changes their parameters.
• Copy Kit to... creates a copy of the selected calibration kit. The cal kit data is stored in the internal
memory and can be exported to a cal kit file using Export Kit.
• Delete Kit removes an imported or user-defined kit from the list of Calibration Kits in Use.
• Export Kit exports calibration kit data to a cal kit file.
• Avail Conn Types adds or deletes connector types.
The Calibration Kits in Use: table is used to select a calibration kit for each connector type. The pull-down
lists contain all kits for the individual connector types.
Remote control:
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:SELect '<ckit_name>'
[S
EN
Se<Ch>:]CORRection:CKIT:SELect '<conn_type>', '<ckit_name>'
Cal Kit Parameter Types
The analyzer uses three types of parameters to describe the calibration standards. The parameter type is
the same for all standards in a kit and therefore annexed to the kit name:
• Universal parameters (no label) describe calibration kit models with highly standardized
components so that the parameters are valid for all calibration kits of the model.
• Typical parameters (labelled typical) approximately describe a calibration kit model. To correct for
deviations between the standards, each kit of the model is individually measured and delivered
with an additional, kit-specific parameter set. Therefore each typical parameter set <kit_name>
typical is complemented by an additional parameter set <kit_name> containing optimized
parameters for an individual kit.
• Ideal parameters (labelled Ideal Kit) describe an idealized calibration kit for each connector type;