R&S ZVL GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
241
The Calibration Kits dialog contains a series of buttons, each of them opening a dialog:
Import Kit imports a cal kit file containing the parameters for a new calibration kit. It is possible to
load several kits, so the analyzer asks whether to set the imported calibration kit active.
Add Kit combines different standards to form a new calibration kit.
View/ Modify Kit adds or deletes standards in an imported or user-defined calibration kit and/or
changes their parameters.
Copy Kit to... creates a copy of the selected calibration kit. The cal kit data is stored in the internal
memory and can be exported to a cal kit file using Export Kit.
Delete Kit removes an imported or user-defined kit from the list of Calibration Kits in Use.
Export Kit exports calibration kit data to a cal kit file.
Avail Conn Types adds or deletes connector types.
The Calibration Kits in Use: table is used to select a calibration kit for each connector type. The pull-down
lists contain all kits for the individual connector types.
Remote control:
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:SELect '<ckit_name>'
[S
EN
Se<Ch>:]CORRection:CKIT:SELect '<conn_type>', '<ckit_name>'
Cal Kit Parameter Types
The analyzer uses three types of parameters to describe the calibration standards. The parameter type is
the same for all standards in a kit and therefore annexed to the kit name:
Universal parameters (no label) describe calibration kit models with highly standardized
components so that the parameters are valid for all calibration kits of the model.
Typical parameters (labelled typical) approximately describe a calibration kit model. To correct for
deviations between the standards, each kit of the model is individually measured and delivered
with an additional, kit-specific parameter set. Therefore each typical parameter set <kit_name>
typical is complemented by an additional parameter set <kit_name> containing optimized
parameters for an individual kit.
Ideal parameters (labelled Ideal Kit) describe an idealized calibration kit for each connector type;