R&S ZVL GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
244
open the View / Modify Standard in <kit_name> dialog check the standard parameters. All other
cont
r
ols are inactive.
In its upper part the Add or View / Modify Calibration Kit dialog contains several controls to do the
following:
• Select a predefined or user-defined Connector Type.
• Insert the (unique) Name of the new kit and assign a Label.
• Qualify whether or not the Offset and Load parameters for the standards are defined and
displayed in Agilent Mode.
Assigning a label to user-defined calibration kits is optional. However, the label is displayed in many
dialogs and can provide useful information about the kit, e.g. its serial number.
The table in the center of the dialog lists all standards in the new kit together with either their gender or
their port assignment (see Restrict Port Assignment) and the label (if defined). Clicking a standard opens
the View / Modify Standard in <kit_name> dialog where it is possible to change the parameters. Four
buttons can be used to change the entries in the list.
• Copy Standards from... opens a dialog to copy standards from another calibration kit to the new
kit.
• Add Standard... selects, defines and adds a new standard to the kit. A calibration kit may only
contain a single standard of each type.
• View / Modify Standard... displays or changes the properties of the selected standard.
• Delete Standard removes the selected standard from the list and from the calibration kit.
Use Copy Standards from... to avoid the (re-)definition of their properties. Standards are copied
together with their parameters and their label, which you can modify once they belong to the new kit.
Copy Standards from
Copies standards together with their parameters from a calibration kit in use to a new calibration kit. This
dialog is opened from the View Calibration Kit dialog (Copy Standards from... button).