R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
354
*RST value
SCPI,
Command
Types
Device-specific, command or query.
Example:
*RST; CALC:DLIN 10
Define the position of the horizontal line in the default dB Mag diagram at +10
dBm.
CALC:DLIN:STAT ON
Display the defined horizontal line.
CALCulate<Chn>:DLINe:STATe <Boolean>
Switches the horizontal line on or off.
<Chn>
Channel number used to identify the active trace.
<Boolean>
ON | OFF - Horizontal line on or off.
*RST value OFF
SCPI, Command Types
Device-specific, command or query.
Example:
See CALC:DLIN.
CALCulate<Chn>:FORMat...
This subsystem determines a post-processing of the measured data in order to obtain various display
formats.
CALCulate<Chn>:FORMat MLINear | MLOGarithmic | PHASe | UPHase | POLar |
SMITh | ISMith | GDELay | REAL | IMAGinary | SWR | COMPlex | MAGNitude
Defines how the measured result at any sweep point is post-processed and presented in the graphical
display.
Note: The analyzer allows arbitrary combinations of display formats and measured quantities; see Trace -
Format and CALCulate<Chn>:PARameter commands. Nevertheless, it is advisable to check which
display formats are generally appropriate for an analysis of a particular measured quantity; see
Measured Quantities and Display Formats.
<Chn>
Channel number used to identify the active trace. If unspecified the numeric
suffix is set to 1.
Parameters
See list of parameters below.
*RST value MLOGarithmic
SCPI, Command
Types
Confirmed, command or query
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'