CALCulate<Chn>:FILTer[:GATE]:TIME:SHOW <Boolean>
Enables or disables permanent display of the gate limits.
<Chn>
Channel number used to identify the active trace
<Boolean>
ON – Time gate permanently displayed
OFF – Time gate hidden
*RST value OFF [–]
SCPI, Command Types
Device-specific, command or query
Example:
See CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN <span>
Defines the span of the time gate.
<Chn>
Channel number used to identify the active trace
<span>
Span of the time gate.
Range [def. unit] 2E-012 s to 200 s. The increment (parameters UP or DOWN) is 0.1 ns. [s]
*RST value 5E-009 s
SCPI, Command Types
Confirmed, command or query
Example:
See CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer
CALCulate<Chn>:FILTer[:GATE]:TIME:STARt <start>
Defines the start time of the time gate.
<Chn>
Channel number used to identify the active trace
<start>
Start time of the time gate.
Range [def. unit] –100 s to +99.999999999998 s. The increment (parameters UP or DOWN) is
0.1 ns. [s]
*RST value –1E-009 s
SCPI, Command
Types
Confirmed, command or query
Example:
*RST; CALC:TRAN:TIME:STAT ON; :CALC:FILT:TIME:STAT ON
Reset the instrument and enable the time domain representation and the
time gate.
CALC:FILT:TIME:STAR 0; STOP 10ns; SHOW ON
Set the start time to 0 ns and the stop time to 10 ns and display the time gate
permanently.
Note: If the start frequency entered is greater than the current stop frequency
(CALCulate<Chn>:FILTer[:GATE]:TIME:STOP), the stop frequency is set to the start
frequency plus the minimum frequency span (CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN).