R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
421
<Boolean>
Enables or disables the deembedding function. It is allowed to change the circuit model and its
parameters while deembedding is enabled.
<Ch>
Channel number.
<Ph_pt>
Physical port number
<Boolean>
ON – Deembedding active
OFF – Deembedding inactive
*RST value OFF [–]
SCPI, Command
Types
Device-specific, command or query
Example:
CALC:TRAN:VNET:SEND:DEEM:TND CSL
Select the Serial C, shunt L circuit model for deembedding.
CALC:TRAN:VNET:SEND:DEEM:PAR:R2 CSL, 2.2E+3; CALC:TRAN:VNET:SEND:DEEM ON
Increase the resistance R2 for the Serial C, shunt L circuit model to 2.2 kV and
enable deembedding.
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:TNDefini
tion FIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL
Selects the circuit model for deembedding.
<Ch>
Channel number.
<Ph_pt>
Physical port number
Parameters
Possible circuit models (character data); see circuit models.
*RST value CSL
SCPI, Command Types
Device-specific, command or query.
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding[:STATe]
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:PARameter
s:C<no> CSL | LSC | CSC | SLC | SCL | SCC,<numeric_value>
Specifies the capacitance values C1, C2 in the different circuit models.
<Ch>
Channel number.
<Ph_pt>
Physical port number
<no>
Number of capacitance in circuit model. If unspecified the numeric suffix is set
to 1.
Range [def. unit] 1 to 2 [–] The number of capacitances depends on the selected circuit model.
Parameters
Possible circuit models (character data); see circuit models.
*RST value (see command type description below)