R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
477
Types
Example:
See MMEMory:LOAD:NWANalyzer.
MMEMory:STORe:RIPPle '<trc_name>', '<file_name>'
Saves the ripple limits associated with a specified trace to a ripple limit file. Ripple limit definitions are
created using the CALCulte<Chn>:RIPPle... commands.
'<trc_name>'
Name of an existing trace in the active setup (string parameter) for which a ripple
limit definition exists.
'<file_name>'
String parameter to specify the name and directory of the created ripple limit file.
The default extension (manual control) for ripple limit files is *.ripple, although
other extensions are allowed. If no path is specified the analyzer uses the current
directory, to be queried with MMEMory:CDIRectory?.
*RST value
SCPI,
Command
Types
Device-specific, no query.
Example:
See MMEMory:LOAD:LIMit.
MMEMory:STORe:SEGMent <Ch>, '<file_name>'
Saves the sweep segment definition of a specified channel to a sweep segment file. Sweep segments are
defined using the [SENSe<Ch>:]SEGMent<Seg>... commands.
<Ch>
Channel number.
'<file_name>'
String parameter to specify the name and directory of the created sweep segment
file. The default extension (manual control) for sweep segment files is *.seglist,
although other extensions are allowed. If no path is specified the analyzer uses the
current directory, to be queried with MMEMory:CDIRectory?.
*RST value
SCPI,
Command
Types
Device-specific, no query.
Example:
See MMEMory:LOAD:SEGMent.
MMEMory:STORe:STATe <numeric_value>,'<file_name>'
Stores the configuration data of the current device settings to a specified settings file.
<numeric_value>
1 (the <numeric_value> is used for compatibility with the SCPI standard but is
ignored).
'<file_name>'
String parameter to specify the name and directory of the created settings file.

Table of Contents