R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
529
Range [def. unit] 1 to 999. [–]
*RST value 1
SCPI, Command Types
Confirmed, command or query
Example:
See CALCulate<Chn>:DATA:NSWeep?.
[SENSe<Ch>:]SWEep:DWELl <delay>
Defines the Meas. Delay time for each partial measurement. This command is not available on R&S
ZVL13 network analyzers.
<Ch>
Channel number.
<delay>
Meas. delay before each partial measurement.
Range [def. unit] 0 s to 3.179551E+002 s. [s] Changing the delay leaves the number of points
unchanged but has an impact on the duration of the sweep.
*RST value 0 s
SCPI, Command
Types
Confirmed, command or query
Example:
FUNC "XFR:POW:S12"
Activate a frequency sweep and select the S-parameter S12 as measured
parameter for channel and trace no. 1.
SWE:DWEL 1
Set a delay of 1 s for each partial measurement.
[SENSe<Ch>:]SWEep:POINts <no_points>
Defines the total number of measurement points per sweep (Number of Points).
<Ch>
Channel number 1.
<no_points>
Number of points per sweep.
Range [def. unit] 2 to 4001. [1]
*RST value 201
SCPI, Command
Types
Confirmed, command or query
Example:
FUNC "XFR:POW:S12"
Activate a frequency sweep and select the S-parameter S12 as measured
parameter for channel and trace no. 1.