R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
461
Parameters
Standard type; for a description refer to the table of standard types.
'<stdlabel_name>'
String parameter defining a label for the cal kit data. An empty string means
that no label is defined.
'<file_name>'
String parameter to specify the name and directory of the Touchstone file to be
loaded. A *.s1p file must be used for one-port standards, a *.s2p file for two-
port standards. If no path is specified the analyzer searches the current
directory, to be queried with MMEMory:CDIRectory?.
<port1_no>,
<port2_no>
Port assignment: One port number for one-port standards, two port number for
two-port standards. If the port numbers are omitted, the cal kit data is valid for
all ports.
*RST value
SCPI, Command
Types
Device-specific, no query.
Example:
MMEM:LOAD:CKIT:SDAT 'N 50 Ohm','Default Kit',MOPEN,'Test
data','test.s1p',1
Load the file Test.s1p
from the current directory in order
to define the properties of an Open (m) standard in the
cal kit named Default Kit for the N 50 R connector type.
Assign the label Test data and specify that the data
standard data is only valid for port no. 1.
Connector and cal kit naming conventions
Connector and calibration kit names must be entered as string parameters. The strings contain the
connector and cal kit names used in the Calibration Kits dialog; a V in the name must be replaced by
'Ohm', e.g.:
'NewKit1' denotes the user-defined calibration kit NewKit1.
'N 50 Ohm Ideal Kit' denotes the N 50 V Dummy Kit.
'ZV-Z21 typical' denotes the cal kit ZV-Z21 typical.
MMEMory:LOAD:CMAP '<file_name>'
Loads a color scheme from a specified NWA color scheme file.
'<file_name>'
String parameter to specify the name and directory of the cal kit file to be loaded.
The default extension (manual control) for color scheme files is *.ColorScheme,
although other extensions are allowed.
*RST value
SCPI,
Command
Types
Device-specific, no query.
Example:
MMEM:LOAD:CMAP
'C:\R_S\Instr\user\Nwa\ColorSchemes\Test.ColorScheme'
Load the previously created color scheme file

Table of Contents