Page #113 background imageLoading...
Page #113 background image
Reference Manual
00809-0100-4004, Rev BA
August 2010
A-15
Rosemount 8800D
PERFORMANCE
SPECIFICATIONS
The following performance specifications are for all Rosemount models
except where noted. Digital performance specifications applicable to both
Digital HART and F
OUNDATION fieldbus output.
Flow Accuracy
Includes linearity, hysteresis, and repeatability.
Liquids—for Reynolds Numbers over 20000
Gas and Steam— for Reynolds Numbers over 15,000
NOTE
As the meter Reynolds Numbers decreases below the stated limit to 10,000 the
accuracy error band will increase linearly to +/-2.0%. For Reynolds Numbers down to
5,000 the accuracy error band will increase linearly from +/-2.0% to +/-6.0%.
Table A-17. Saturated Steam Flow Rate Limits (Assumes Steam Quality is 100%)
Process
Pressure
Flow Rate
Limits
Minimum and Maximum Saturated Steam Flow Rates
for line sizes 10 inch/DN 250 through 12 inch/DN 300
10 Inch/DN 250 12 Inch/DN 300
Rosemount 8800D Rosemount 8800DR Rosemount 8800D Rosemount 8800DR
lb/hr kg/hr lb/hr kg/hr lb/hr kg/hr lb/hr kg/hr
15 psig
(1,03 bar G)
max
min
35519
3175
16111
1440
22534
2015
10221
914
50994
4554
23130
2066
35519
3175
16111
1440
25 psig
(1,72 bar G)
max
min
46618
4570
21146
2073
29575
2308
13415
1047
66862
5218
30328
2367
46618
4570
21146
2073
50 psig
(3,45 bar G)
max
min
73748
4575
33452
2075
46787
2903
21222
1317
105774
6562
47978
2976
73748
4575
33452
2075
100 psig
(6,89 bar G)
max
min
126660
5996
57452
2720
80356
3804
36449
1725
181663
8600
82401
3901
126660
5996
57452
2720
150 psig
(10,3 bar G)
max
min
178808
7125
81106
3232
113440
4520
51455
2050
256457
10218
116327
4635
178808
7125
81106
3232
200 psig
(13,8 bar G)
max
min
230722
8092
104654
3670
146375
5134
66395
2329
330915
11607
150101
5265
230722
8092
104654
3670
300 psig
(20,7 bar G)
max
min
334810
9749
151867
4422
212411
6185
96348
2805
480203
13983
217816
6343
334810
9749
151867
4422
400 psig
(27,6 bar G)
max
min
440085
11442
199619
5190
279200
7259
126643
3293
631195
16411
286305
7444
440085
11442
199619
5190
500 psig
(34,5 bar G)
max
min
547165
14226
248190
6453
347134
9025
157457
4094
784775
20404
355968
9255
547165
14226
248190
6453
Digital and Pulse Output
±0.65% of rate
Note: The accuracy for the 8800DR, line sizes 6 to 12 inch
(150 to 300mm), is ±1.0% of rate.
Analog Output
Same as pulse output plus an additional 0.025% of span
Digital and Pulse Output
±1.0% of rate
Note: The accuracy for the 8800DR, line sizes 6 to 12 inch
(150 to 300mm), is ±1.35% of rate.
Analog Output
Same as pulse output plus an additional 0.025% of span
Accuracy limitations for gas and steam:
- for
1
/2- and 1-in. (DN 15 and DN 25):
max velocity of 220 ft/s (67.06 m/s)
- for Dual-style meters (
1
/2-in. to 4 in.):
max velocity of 100 ft/s (30.5 m/s)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rosemount 8800D Series and is the answer not in the manual?

Rosemount 8800D Series Specifications

General IconGeneral
TypeVortex Flow Meter
Measurement PrincipleVortex Shedding
Operating Temperature-40 to 185°F (-40 to 85°C)
Storage Temperature-40 to 185°F (-40 to 85°C)
Communication ProtocolsFOUNDATION Fieldbus
Material316 stainless steel
Fluid TypeLiquid, Gas, Steam
Line Size1/2 to 8 inches (15 to 200 mm)
ApprovalsATEX, IECEx, CSA, FM
Output Signal4-20 mA
Process ConnectionFlanged, Wafer, Threaded
Electrical ConnectionM20
Protection RatingIP66
Mounting StyleWafer, Flanged

Related product manuals