EasyManuals Logo

Rossmax NA100 Service Manual

Rossmax NA100
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
6. Recommended Parts List:
Please refer the below parts to the exploded drawing
Medication cup Mouthpiece Air tube Filter Adult and Child mask Cover and diverter Nozzle
5. Trouble Shooting:
Symptom A Cannot be turned on
Figure
Root causes Defective power key Defective power wires
Correction Replace the power key Replace the power wires
Symptom B Motor operates but no spray is generated
Figure
Root Causes Run out of liquid medication Air tube is not connected properly
Correction Rell the medication cup Re-connect the air tube and
medication cup properly
Symptom C Low spray volume
Figure
Root Cause Valve is open Nozzle is clogged Defective lter
Correction Adjust the valve Rinse the nozzle with
water
Replace the lter
Symptom D Machine shuts down suddenly during operation
Figure
Root Cause Motor automatically shuts down due to overheat
Correction Wait for the temperature to drop, which takes about 20 minutes
www.rossmax.com
Rossmax Swiss GmbH,
Tramstrasse 16, CH-9442 Berneck, Switzerland
INXXXXXXXXXXXXXXX
S&S_ServiceManual_NA100_EN_SW_ver1544
Service Manual
Document No.: SM-11-012 Version: 2
ROSSMAX Model No.
NA100 Nebulizer

Other manuals for Rossmax NA100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rossmax NA100 and is the answer not in the manual?

Rossmax NA100 Specifications

General IconGeneral
Product TypeNebulizer
Compressor TypePiston
Nebulization Rate≥ 0.2 ml/min
Weight1.5 kg
Dimensions164 x 150 x 94.5 mm
Power SupplyAC 230V / 50Hz
Sound Level≤55 dB
Particle Size≤ 5 μm
Medication Capacity5 ml (cc)