EasyManua.ls Logo

SANKO SWT-7000 III - User Manual

Default Icon
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
SANKO Coating Thickness Meter
SWT-7000Ⅲ/7100Ⅲ
Instruction Manual
CAUTIONS:
Before using the Meter, read this INSTRUCTION MANUAL
thoroughly and use the Meter correctly.
Keep this INSTRUCTION MANUAL carefully and refer to this
when necessary.
In the event of any doubt arising, the original
INSTRUCTION MANUAL (Japanese) is to be of final authority.
SANKO ELECTRONIC LABORATORY CO., LTD.
Tokyo・Osaka・Sendai・Nagoya・Fukuoka・Kawasaki
December 2014

Questions and Answers

  • K
    Kristin SolisAug 1, 2025
    What to do if SANKO SWT-7000 III Measuring Instruments display 'BATTERY is dead! Replace all of them with NEW BATTERY. ?Power ???? ERROR?'?
    • L
      Lauren BrownAug 2, 2025
      If your SANKO Measuring Instruments show this message, the issue is likely due to low batteries or starting to press probe to object soon after switching on. Replace the batteries with new ones and hold the probe in the air, away from objects and metals, while the message is displayed.
  • J
    Jordan DavisAug 7, 2025
    What to do if SANKO SWT-7000 III show 'ERROR? Connect a probe before switched on. ?Power ????'?
    • R
      Rose MunozAug 7, 2025
      This message appears when the ON/OFF key is pressed before connecting the probe. Ensure the probe is connected before pressing the ON/OFF key.
  • J
    Jessica LucasAug 11, 2025
    Why SANKO Measuring Instruments have no response upon pressing the ON/OFF key?
    • R
      Rodney JohnsonAug 11, 2025
      If your SANKO Measuring Instruments do not respond when you press the ON/OFF key, the batteries may be worn out. Replace them with two new batteries.
  • S
    Sydney PooleAug 16, 2025
    What to do if SANKO SWT-7000 III are unable to transfer data?
    • R
      rjacksonAug 17, 2025
      If you're having trouble transferring data from your SANKO Measuring Instruments, it could be due to a faulty USB cable connection or broken wiring, or the PC may not be ready. Make sure the cable is firmly connected. Also: 1. Install the attached CD correctly. 2. Set the Comport No. correctly.
  • D
    David EspinozaAug 21, 2025
    What to do if SANKO SWT-7000 III Measuring Instruments batteries have worn out?
    • C
      catherinejordanAug 21, 2025
      If the batteries in your SANKO Measuring Instruments have worn out, replace them with new ones.
  • W
    wgilbertAug 27, 2025
    What to do if SANKO SWT-7000 III experience sudden interruption of data transfer?
    • W
      wsteeleAug 27, 2025
      If data transfer is suddenly interrupted, there might be an issue with the PC, the main unit, or the cable. Check the PC for any problems.

Summary

Safety and Handling Precautions

General Safety Warnings

Covers critical safety warnings regarding water, contamination, handling, and tampering.

Battery and AC Adapter Safety

Details safety precautions for battery installation, storage, and AC adapter usage.

General Handling and Storage Safety

Advises on safe handling, storage conditions, and avoiding material contact.

General Usage Notes

Provides essential notes on manual reading, careful handling, and maintenance.

Getting Started

Contents in Package

Lists all items included in the product package for user verification.

Device Overview and Components

Component Identification

Identifies and describes key components of the coating thickness meter.

Battery Installation

Step-by-step guide for inserting batteries and related safety precautions.

Basic Operations

Probe Connection and Disconnection

Instructions for attaching and detaching the measurement probe safely.

Probe Handling and Application

Guidance on proper probe grip and application for accurate measurements.

How to Operate the Device

Switching Power Source On and Off

Instructions for switching the device's power source on and off.

Zeroing Procedure

Step-by-step guide for calibrating the meter to a zero reference point.

Calibration Standard (CAL)

Procedure for calibrating using a standard thickness foil for accurate readings.

Special Case Zeroing (Multi-layers)

Method for zeroing the meter on multi-layered coatings for specific measurements.

2-Foil Calibration Method

Procedure for 2-foil calibration when standard zeroing is difficult.

Deleting Calibration Data

Instructions for erasing stored calibration data from the device.

Measuring

Measuring Procedure Overview

General guidance and steps for performing measurements with the device.

Function Switching

Non-Interrupt Measurement Mode

Enables continuous measurement by sliding the probe along the surface.

Switching Measurement Resolution

Adjusts measurement resolution for precise thickness inspection up to 500 μm.

Auto-Power-Off Setting

Configures automatic power-off to save battery life after periods of inactivity.

Substrate Mode Switching (FN-325 Probe)

Switches between AUTO, Fe, and NFe modes for FN-325 probe compatibility.

Backlight ON;OFF Control

Enables or disables the display backlight for visibility in dark conditions.

Transferring Data (SWT-7100 III)

PC Data Transfer Procedure

Steps for transferring measured data to a PC using a USB cable.

Improving Measuring Accuracy

Factors Affecting Measuring Accuracy

Details factors like zero plate, thickness standards, and environmental conditions.

Troubleshooting

Common Device Issues and Solutions

Addresses common symptoms like no response, battery errors, and connection problems.

Probe and Data Transfer Troubleshooting

Resolves issues related to probe malfunctions and data transfer interruptions.

Specifications

Unit Specifications

Details model names, display, ranges, functions, power, and dimensions of the unit.

Probe Specifications (FN-325 Model)

Lists specifications for the FN-325 probe, including ranges, accuracies, and accessories.

Probe Specifications (Other Models)

Details specifications for various other probe models like Fe-2.5, Fe-10, Fe-20.

Probe Specifications (Fe-0.6 Pen, NFe Models)

Lists specifications for Fe-0.6Pen and NFe series probes, including ranges and accuracies.

Reference: Principle of Measurements

Measurement Principles Explained

Explains the Electro-Magnetic and Eddy Current principles used in measurements.

Manufacturer Information

Sanko Electronic Laboratory Co., Ltd. Contacts

Provides contact details and branch information for the manufacturer.

SANKO SWT-7000 III Specifications

General IconGeneral
BrandSANKO
ModelSWT-7000 III
CategoryMeasuring Instruments
LanguageEnglish