18
Further Options
Print Data
If you have connected an appropri-
ate printer to the printer port (1) on
your arium system, you have two
print options:
– Print Interval Option
With the Print Interval Option, the
system can be set up to print data
at regular intervals when in Normal
Operation.
– Print Screen Option
The Print Screen Option will only
print the last data recorded.
Note
The arium system has a 9-pin
connection for a serial printer. The
printer must use a separate power
supply and be configured as fol-
lows: 19,200 baud, 8-bit, no parity.
To select a print option:
• Press MENU to get into the Main
Menu:
• Select “Utilities” and press ENTER
to confirm.
• Select “Print” and press ENTER
to confirm.
• Select the desired option and press
ENTER to confirm.
– With the Print Interval Option, the
system can be set up to print data
at regular intervals. If you have
selected “Print Interval”, you can
use the control keys for the cursor
to set the desired length of interval
in minutes when prompted by the
display.
– “Print Screen”
The Print Screen Option will only
print the last data recorded. You can
generate a printout of the screen
by pressing ENTER once. After
printing, the system returns to the
previously activated Operate Mode.
PRINT
Print Interval
Print Screen
UTILITIES
Print
Timer Contr.
MENU Utilities
Maintenance
Setup
Pump Protection
If you are using a separate tank
for the feed water supply, you can
connect the water tank and the
arium unit via port (2) with a pump
protection for tank feed systems.
When the tank is empty, the pump
is turned off and water will not
recirculate (only with arium tank).
Printer Port (1) and pump
interlock (2)
2
1