Calibration of the System
The system can be calibrated using
an optional N.I.S.T. calibration
module available from Sartorius.
Proceed as follows:
Turn the power switch on the arium
System to off (“O”) and disconnect
power cord from the unit.
Disconnect the inlet water line from
the arium system. Open outlet valve
to depressurize the system.
Open the front doors and move the
red cartridge pack #2 to the side by
rotating the pack to the right.
Pull up the yellow sealing hood
from the gray resistivity cell cable.
Disconnect the cell from the
gray cable. Attach the calibration
module where the cell cable was
connected.
Close the front doors, reattach
the power cord and turn the power
switch on (“I”) – wait for the unit
to complete the system check.
Caution!
Danger of water spillage.
Press MENU to get into the Main
Menu.
Use the cursor to select “Mainte-
nance”. Press ENTER to confirm.
Use the DOWN arrow y to
select “Calibration”. Press ENTER
to confirm.
After pressing ENTER, the
calibration will begin.
See Owners Manual
for Details.
Push ENTER to Start
MAINTENANCE
UV Life Timer
Sanitization
MENU Utilities
Maintenance
Setup
21
This will take a few minutes
to complete:
When calibration is successful, the
display will show the following:
Turn the power switch on the arium
System to off (“O”) and remove the
power cord.
Open the front doors and unplug
the calibration module from the
cell.
Reattach the gray cable to the cell.
The yellow sealing hood should
form a tight seal around the
connection.
Close the doors and reattach the
power cord.
Connect the inlet water line and
power on the system (“I”).
The calibration is now complete.
The unit is ready for normal
operation.
Calibration
Finished
Calibration
In Process
!