Calibration of the System
The system can be calibrated using
an optional N.I.S.T. calibration
module available from Sartorius.
Proceed as follows:
• Turn the power switch on the arium
System to off (“O”) and disconnect
power cord from the unit.
• Disconnect the inlet water line from
the arium system. Open outlet valve
to depressurize the system.
• Open the front doors and move the
red cartridge pack #2 to the side by
rotating the pack to the right.
• Pull up the yellow sealing hood
from the gray resistivity cell cable.
• Disconnect the cell from the
gray cable. Attach the calibration
module where the cell cable was
connected.
• Close the front doors, reattach
the power cord and turn the power
switch on (“I”) – wait for the unit
to complete the system check.
Caution!
Danger of water spillage.
• Press MENU to get into the Main
Menu.
• Use the cursor to select “Mainte-
nance”. Press ENTER to confirm.
• Use the DOWN arrow y to
select “Calibration”. Press ENTER
to confirm.
• After pressing ENTER, the
calibration will begin.
See Owners Manual
for Details.
Push ENTER to Start
MAINTENANCE
UV Life Timer
Sanitization
MENU Utilities
Maintenance
Setup
21
This will take a few minutes
to complete:
When calibration is successful, the
display will show the following:
• Turn the power switch on the arium
System to off (“O”) and remove the
power cord.
• Open the front doors and unplug
the calibration module from the
cell.
• Reattach the gray cable to the cell.
• The yellow sealing hood should
form a tight seal around the
connection.
• Close the doors and reattach the
power cord.
• Connect the inlet water line and
power on the system (“I”).
The calibration is now complete.
The unit is ready for normal
operation.
Calibration
Finished
Calibration
In Process
!