23
System Inactive Mode
If the Operate|Stop button is
pressed while the system is on Nor-
mal Operation the arium 611VF is
deactivated. This mode is intended
purely for troubleshooting and
should not be used for prolonged
periods of time. Otherwise the
entire system will have to be
flushed and sanitized or cartridge
packs replaced.
When the system is inactive, the
display shows:
If you need to shut down arium
611VF for an extended period of
time, you should completely drain
it, and remove all cartridge packs
and the ultrafilter to prevent the
growth of bacteria. If the system
has remained inactive and full
of water for more than 96 hours,
you must drain the unit, install
new cartridges and sanitize the
unit prior to use.
System Inactive
Open product outlet valve
Push the draw-off lever forwards to close the product outlet valve.