Fuse Replacement
arium 611VF has two fuses for
the unit.
Warning!
Severe electrical shock hazard
or danger of electrocution!
Only qualified personnel may
service and repair arium 611VF.
Disconnect arium 611VF from the
AC power outlet before mainte-
nance and servicing.
Ensure all tubing connections are
tight to avoid leakage.
Depressurize the unit by pushing
back the draw-off lever to the right
before opening the front doors to
exchange the cartridges.
Turn off arium 611VF using
the main power switch (“O”)
and disconnect it from the AC
power outlet.
The fuse drawer with two safety
fuses is located on the left side of
the housing above the main power
switch.
Gently press the locking device of
the fuse drawer upwards and pull
out the fuse drawer (see figure).
Caution!
Danger of damage to components!
First determine why the fuse blew
and eliminate the cause. Then
replace any defective fuses with
new ones.
Remove the old fuses and replace
with fuses of the same type and
rating.
Place fuse drawer back into the
housing until it clicks into place.
Reconnect the unit to the AC power
outlet and turn the unit back on
using the main power switch (“I”).
The unit is now ready to operate
again.