Care and Maintenance
Operating Instructions Arium
®
Comfort I, Comfort II 59
t Start the rinsing process.
y Flush out the alkaline cleaning solution with a ten-minute flushing and
rinsing process.
You can cancel the rinsing process “
Cancel
” via the (L) key.
Prepare acid cleaning solution. Rinse the alkaline solution from the

recommended) and add the cleaning solution “Acid Cleaner Solution 2”.
t Attach the “Tank Outlet” tubing (4), and the rinsing (7) and cleaning (3)
tubing to the container with the acid cleaning solution (8).
t Start cleaning.
r r System cleaning
Start rinsing process
Start
r r System cleaning
Rinsing
10 min
Cancel
r r System cleaning
Prepare acid cleaning solution
according to operating
instructions
OK
r r System cleaning
Attach the “Tank outlet” tubing
and the rinsing and cleaning
tubing to the container with
the cleaning solution.
OK
r r System cleaning
Start cleaning
Start