Care and Maintenance
60 Operating Instructions Arium
®
Comfort I, Comfort II
y Rinse the Arium
®
Comfort system with the acid solution for 30 minutes.
You can cancel the cleaning process using the (L) “
Cancel
” key.
Information and Instructions on Disposal “Acid Cleaner Solution 2”
This solution can generally be emptied down the drain after the cleaning
process; however, please check with your local authorities first before doing
so to ensure compliance.
t Attach the “Tank Outlet” tubing and the rinsing and cleaning tubing to
the drain.
t Start the rinsing process.
y The cleaning solution is flushed out. The rinsing takes 30 minutes.
t Connect the “Tank Outlet“ tubing (4) to the Arium
®
Bagtank (5).
t Remove the dummy plug from the softener cartridge (1). Remove the
cleaning tubing (3) including the ¼” - ¼” adapter (2) and insert the open
end of the tubing into the softener cartridge (1). The rinsing tubing
should be left in the drain (7)!
The system cleaning timer is reset automatically.
r r System cleaning
Cleaning
30 min
Cancel
r r System cleaning
Direct “Tank outlet” tubing and
rinsing and cleaning tubing
towards the drain
OK
r r System cleaning
Start rinsing process
Start
r r System cleaning
Rinsing
30 min
Cancel
r r System cleaning
Connect “Tank outlet” tubing
to tank and remove cleaning
tubing
OK