Care and Maintenance
Operating Instructions Arium
®
Comfort I, Comfort II 61
9.2 Start Bag Replacement
Select the menu item Care Start bag replacement to replace the plastic
bag in the Arium
®
Bagtank. The bag should be replaced every 6 months.
t Close the tank valve (ball cock) and disconnect the “Tank Outlet” tubing
from the Arium
®
Comfort system. Attach the open end of the tubing to
the drain.
t Open the tank valve and empty the tank.
Use an Arium
®
Bagtank 50 or 100 with built-in pump; alternatively the bag
can be emptied using the Tank Kit 2 (OUT). This way, the Arium
®
Bagtank
can be emptied faster. Follow the installation instructions for the Arium
®
Bagtank.
t Turn off the Arium
®
Bagtank at the power switch (Bagtank 50 and | or
100 built-in pump only). Replace the bag according to the operating
instructions.
t Reconnect the “Tank Outlet“ tubing to the Arium
®
Comfort system.
t Turn on the Arium
®
Bagtank at the power switch. The bag replacement is
complete.
With TOC critical applications, Sartorius recommends rinsing the bag after
the bag has been replaced (see Chapter "9.6 Tank Rinsing", page 70).
r r Bag replacement
Close tank valve, disconnect
“Tank outlet” tubing from
device and direct towards
drain
Cancel OK
Last chance to cancel
r r Bag replacement
Open the tank valve to empty
the tank
OK
r r Bag replacement
Replace the bag according to
the operating instructions
OK
r r Bag replacement
Connect “Tank Outlet” tubing
to device
OK