Care and Maintenance
Operating Instructions Arium
®
Comfort I, Comfort II 77
Pos. Description
1 Quick connector
2 Vent valve
3 Bell assembly
t Attach the bell assembly to the final filter.
t Press the new final filter into the quick connector of the
display | dispenser unit.
t To rinse and to vent the final filter: Remove the specified amount of
water (water dispensing volume, see instructions for the Sterile Plus or
Cell Plus final filter), e.g. via manual or volume-controlled water
dispensing.
t If a Sterile Plus (sterile filter) final filter is used: Open the vent valve.
t After dispensing ultrapure water, attach the protective cap to the bell
assembly.
t If a Sterile Plus (sterile filter) final filter is used: Close the vent valve.
If a newly installed final filter clogs rapidly after installation, your Arium
®
system may need to be sanitized to remove bacterial contaminants (see
Chapter "9.1 System Cleaning", page 56). After replacement, install a
new final filter.
t After you have replaced the sterile final filter, the timer must be reset
manually. To reset the timer, select “Final Filter Timer” in the Care main
menu.
t Confirm the process.
y The timer is now reset and the system will then switch to the operating
mode again.
The warning to replace the final filter will not be deactivated until the timer
is reset.
1
2
3
Care
Backflushing
UV lamp timer
Replace cartridge
Final filter timer
r Final filter timer
Reset timer
Settings
Back