Care and Maintenance
80 Operating Instructions Arium
®
Comfort I, Comfort II
Restart after depressurization
t Next, switch your Arium
®
Comfort system back on via the power switch.
The system starts up. Next, connect the tubing for device operation
according to the operating instructions.
t Place a vessel (1l) under the water outlet and start the rinsing process.
y The system is rinsed for one minute. Your Arium
®
then switches to the
operating mode.
If the tank level falls to < 2 L during the rinsing process, the rinsing is inter-
rupted and the system waits for the tank to fill to a sufficient level (> 5 L).
Afterwards, the rinsing process is resumed.
9.14 Replacing Electrical Fuses
The Arium
®
Comfort system has two fuses.
They are located in a fuse drawer on the left hand side over the main power
switch (see Chapter "3.3 Arium
®
Comfort Electrical Connections",
page 11).
WARNING
Only trained and qualified personnel may replace defective fuses.
WARNING
Always disconnect the system from the AC power outlet before replacing
defective electrical fuses.
WARNING
Remove the old fuses from their sockets and ensure that they are always
replaced with fuses of the same type and rating.
t Gently press the locking device of the fuse drawer towards the left and
pull out the fuse drawer.
t Replace the defective fuses.
t Push the fuse drawer back into the housing until it clicks into place.
Care Rinsing
Connect the tubing for device
operation according to the
operating instructions
OK
Care Rinsing
Place vessel (1l) under
the dispenser and start
dispensing water
Start
Care Rinsing
Rinsing
1 min
Cancel