Technical Specifications
88 Operating Instructions Arium
®
Comfort I, Comfort II
Feed Water Requirements
Exclusively potable tap water (feed water must meet drinking water standards of the USA, European Union and

3
): 360 ppm
TOC: < 2000 ppb
Colloid index or SDI: < 5
Free chlorine: < 4 ppm
Turbidity: < 1 NTU
Iron (evaluated as total FE content): < 0.1 ppm
Feed water pH: 4 – 10
Temperature range: 2 °C – 30 °C
Pressure
1)
: min. 0.5 bar, max. 6.9 bar
(A feed water input pressure of > 2 bar is recommended for optimal system
operation. If the input pressure is < 2 bar, this will lower the flow performance
in L/h of the system).
1)
Dynamic pressure | flow pressure 100 L/h