Operating Instructions BioPAT
®
Xgas 37
Technical Data
8 Technical Data
O
2
Sensor Unit
Gas O
2
Concentration range 1 50% by volume, others available upon request
Measuring principle Zirconium dioxide
Accuracy < 0.2% FS2 ± 3% Rdg2 on value
Drift < ± 2% value | year
Lifetime of sensor
element
Approx. 15,000 operating hours | approx.
2 years
Temperature inside the
sensor unit
580°C
CO
2
Sensor Unit
Gas CO
2
Concentration range 0 10% by volume, others available upon request
Measuring principle Infrared
Accuracy < 0.2% FS2 ± 3% Rdg2 on value
Drift < ± 2% value | year
Lifetime of sensor
element
Approx. 25,000 operating hours | approx.
3 years
Temperature inside the
sensor unit
3°C higher than process temperature

Table of Contents