Cubis
®
MCE Operating Instructions 29
System Settings
7.3 Parameter List
7.3.1 Parameters in the “Setup” Main Menu
Parameters in the “Balance” Submenu
Parameters Setting values Explanation
AMBIENT V.STABLE Sets the ambient conditions to “very stable”: Activates a fast change in the weight values
in the event of a load change with a high output rate.
Recommended for the following work environment:
Very stable table near the wall
Closed and calm room
STABLE* Sets the ambient conditions to “stable”.
Recommended for the following work environment:
Stable table
Slight movement in the room
Slight draft
UNSTABL. Sets the ambient conditions to “unstable”: Activates the delayed change in weight values
with a reduced output rate.
Recommended for the following work environment:
Simple office desk
Room with moving machinery or personnel
Slight air movement
V.UNSTBL. Sets the ambient conditions to “very unstable”: Activates a significantly delayed change
in the weight values and long wait for stability with a further reduction in the output
rate.
Recommended for the following work environment:
Noticeable and slow floor vibrations
Noticeable building vibrations
Weighed goods moved
Very strong air movements
APP FILT. FINAL.RD.* Activates a filter that enables a fast change in the display for very fast load changes.
Display changes with minimal load changes (in the digit range) occur more slowly.
FILLING Activates a filter that enables a very fast change in the display with minimal load
changes (e.g. when filling containers).
REDUC. Activates a weak but fast filter that always behaves in the same way for load changes
(e.g. when filling automated systems).
OFF Deactivates the active application filter.
STABIL. MAX ACC. Sets the stability to “maximum accuracy”.
V. ACC. Sets the stability to “very accurate”.
ACC.* Sets the stability to “accurate”.
FAST Sets the stability to “fast”.
V. FAST Sets the stability to “very fast”.
MAX.SPEED Sets the stability to “maximum speed”.
* Factory setting

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCE Series and is the answer not in the manual?

Sartorius Cubis MCE Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCE Series
CategoryScales
LanguageEnglish

Related product manuals