44 Cubis
®
MCE Operating Instructions
Operation
Calibration and Subsequent Adjustment
Requirements
The weighing pan is unloaded.
The operating display shows a stable weight value.
The device is not locked against external calibration.
Keep calibration weight ready for use.
Procedure
t To set the external calibration in the menu: In the “SETUP” / “BALANCE” menu,
for the “CAL./ADJ.” parameter, select external calibration with preset calibration
weight or external calibration with user-defined calibration weight as described
above.
t To set the sequence of the adjustment function to calibration and subsequent
automatic adjustment: In the “SETUP” / “BALANCE” menu, for the “CAL.SEQ.”
parameter, select the “ADJUST” configuration value.
t Exit the menu.
t Press the [Zero] key.
t Press the [Adjust] button.
t Place the indicated calibration weight on the balance. The device is automatically
adjusted after calibration, if the calibration weight placed on the device is within
the specified limits.
y The [+] sign is shown if the applied weight is too high.
y The [-] sign is shown if the applied weight is too low.
y The [CAL.END] message indicates the end of calibration.
t Remove the calibration weight after calibration is complete.
Manually Starting Adjustment after Calibration
Requirements
The weighing pan is unloaded.
The operating display shows a stable weight value.
The device is not locked against external calibration.
Procedure
t To set the external calibration in the menu: In the “SETUP” / “BALANCE”” menu,
for the “CAL./ADJ.” parameter, select external calibration with preset calibration
weight or external calibration with user-defined calibration weight as described
above.
t To set the sequence of the adjustment function to calibration without
subsequent automatic adjustment: In the “SETUP” / “BALANCE” menu, for the
“CAL.SEQ.” parameter, select the “CAL/ADJUST” configuration value.
t Exit the menu.
t Press the [Zero] key.
t Press the [Adjust] button.
t Place the indicated calibration weight on the balance. The calibration can only be
started, if the calibration weight used is within the defined limits.
y The [+] sign is shown if the applied weight is too high.
y The [-] sign is shown if the applied weight is too low.
t Press the [Confirm] button on the keypad. This starts the calibration.
y The [CAL.END] display indicates the end of calibration.
t Remove the calibration weight after calibration is complete.