Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential 29
Startup
t Take the device out of the packaging and place it at the intended
installation location.
t Remove the left side cover:
t Reach behind the side cover and pull the side cover to the side (1).
t Pull the side cover up and off to remove it (2).
t Remove the following parts from the inside of the device:
− Tubing
− Tubing adapter
− Hose release tool
− Power cord
6 Startup
We recommend having the initial startup of the device carried out by
Sartorius Service. Please contact Sartorius Service in this regard.
6.1 Connecting the AC Adapter
CAUTION
Risk of electric shock due to incorrect handling of power cables!
The use of damaged or non-standard power cables as well as the mishandling of
power cords can cause fatal electric shock or equipment damage.
t Only connect the device (protection class 1) to properly installed power sockets
with protective grounding conductors (PE) with a fuse of a maximum of 16 A.
t Use only standard cables that have protective grounding conductors for
operation.
t Never disconnect the device from the protective grounding conductor.
t Connect to the power supply according to the regulations of your specific
country.
t Never plug the power cable into the mains wall outlet when it is disconnected
from the device.
t Make sure that the power plug or another suitable disconnecting device for
the power can be easily reached in case of danger.
NOTICE
Equipment damage due to operation with third-party equipment!
The use of third-party power supplies not authorized by Sartorius may cause
damage to the device.
t Use only the original Sartorius power supply.
Procedure
t Check whether the plug design of the power cord is equivalent to your
country’s standard.
t If required: Call Sartorius Service or your dealer.
2
1