Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential 31
Startup
6.2 Starting up the Device
6.2.1 Startup Overview
When the system check is complete, the dialog box “Language” appears.
The wizard automatically performs all startup steps. Startup takes approx. 120 minutes
and cannot be canceled.
All system settings (e.g. date, time, displayed values) that are made during the
startup can be changed afterwards in the “Settings” menu (see Chapter “7.6
Changing System Settings,” page 51).
Startup includes the following tasks:
Startup procedure Chapter, page
Set the language 6.2.2, page 31
Start startup mode 6.2.3, page 32
Set date and time 6.2.4, page 32
Set the displayed values 6.2.5, page 33
Insert pretreatment cartridge (only arium
®
mini plus) 6.3, page 34
Insert ultrapure water cartridge 6.4, page 35
Insert bag (only arium
®
mini plus or arium
®
mini) 6.5, page 36
Connect tubing 6.6, page 38
Rinse pretreatment cartridge (only arium
®
mini plus) 6.7, page 40
Rinse the ultrapure water cartridge 6.8, page 41
Connect final filter 6.9, page 44
Rinse final filter 6.10, page 44
6.2.2 Setting the Language
Procedure
t Select the language for the display. The factory setting is English.
t Confirm the selection with the [l] button.
y The dialog box “Startup” appears.
Language
English
Deutsch
Français
Español