34 Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential
Startup
6.3 Inserting Pretreatment Cartridge (only arium
®
mini plus)
A pretreatment cartridge must be inserted in the device. The cartridge holder for the
pretreatment cartridge is marked “R”.
Procedure
t Remove the front cover.
t Confirm removal of the cover with the [OK] button.
y The dialog box “Insert cartridge” appears.
t NOTICE Equipment damage due to impurities! If dirt or foreign objects enter the
ultrapure water circulation, individual device components can become clogged or
wear out faster. Do not touch the connections of the pretreatment cartridge.
This will prevent impurities from getting into the device.
t Remove the plugs from all three connections of the pretreatment cartridge.
t Hold the pretreatment cartridge in the upper area and insert it into the holder
marked “R”. To do this, push the pretreatment cartridge straight into the guide
rails of the cartridge holder.
t Press the pretreatment cartridge firmly into the cartridge holder until it locks
into the guide rails with a distinct clicking sound on both the left and right sides.
Check that the pretreatment cartridge has been installed securely by turning it
slightly.
t Confirm the insertion of the pretreatment cartridge with the [OK] button.
y The “Insert cartridge” dialog box (Step 2) appears. An ultrapure water cartridge
must be inserted (see Chapter 6.4, page 35).
1
Insert cartridge
Insert the pretreatment cartridge (R) as
described in the manual.
OK
1
Insert cartridge
Insert the ultrapure water cartridge (R) as
described in the manual.
OK