6.4 Inserting Ultrapure Water Cartridge
An ultrapure water cartridge must be inserted in the device. The cartridge holder for
the ultrapure water cartridge is marked “L”.
Procedure
t Remove the front cover.
t Confirm removal of the cover with the [OK] button.
y The dialog box “Insert cartridge” appears.
t NOTICE Equipment damage due to impurities! If dirt or foreign objects enter the
ultrapure water circulation, individual device components can become clogged or
wear out faster. Do not touch the connections of the ultrapure water cartridge.
This will prevent impurities from getting into the device.
t Remove the plugs from the two external ultrapure water cartridge connections.
t Hold the ultrapure water cartridge in the upper area and insert it into the holder
marked “L”. To do this, push the pretreatment cartridge straight into the guide
rails of the cartridge holder.
t Press the ultrapure water cartridge firmly into the cartridge holder until it locks
into the guide rails with a distinct clicking sound on both the left and right sides.
Check that the ultrapure water cartridge has been installed securely by turning it
slightly.
t Confirm the insertion of the ultrapure water cartridge with the [OK] button.
y The dialog box “Attaching the cover” appears.
t Put the front cover on the device.
t Confirm that the cover has been put on with the [OK] button.
1
Insert cartridge
Insert the ultrapure water cartridge (L) as
described in the manual.
OK
1
Put the cover back on
Please put the front cover back on.