38 Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential
Startup
6.6 Connecting Tubing
6.6.1 Connecting Feed Water Tubing (only arium
®
mini plus or arium
®
mini)
To prevent the entry of particles into the device, it is recommended to use the
supplied screens.
Procedure
t Connect the feed water tubing to the connection labeled “Inlet” on the rear of
the device.
t NOTICE Water leakage due to excessive inlet pressure! If the water pressure is
too high: Water can leak out of the device. Check the water pressure inlet
specification (see Chapter “14.6 Feed Water Quality,” page 75).
t If required: Reduce the inlet pressure.
t NOTICE Water leakage can occur if feed water tubing is not watertight! If the
feed water tubing is deformed or not inserted deep enough: Water can leak out.
After startup, make sure that all external water connections are leak proof.
t Connect the feed water tubing to the feed water supply.
6.6.2 Connecting Tank Filling Tubing (only arium
®
mini)
To prevent the entry of particles into the device, it is recommended to use the
supplied screens.
NOTICE
Contamination of tank filling tubing!
If the tank filling tubing is improperly stored: The tank filling tubing can be
contaminated. Sufficient quality of the pure water that the bag is filled with is no
longer guaranteed in this case.
t Do not leave the tank filling tubing on the floor or near sources of
contamination (e.g. in the vicinity of an outlet).
t If the tank filling tubing cannot be kept secure while connected to the device:
Remove the tank filling tubing and keep it in a clean place.
t If the tank filling tubing is contaminated:
Replace the bag (see Chapter 8.5.2, page 56).
Replace the tank filling tubing.
t Use the tank filling tubing exclusively for filling.
Procedure
t Connect the tank filling tubing to the connection labeled “Inlet” on the rear
of the device.
t NOTICE Water leakage due to excessive inlet pressure! If the water pressure is
too high: Water can leak out of the device. Check the water pressure inlet
specification (see Chapter “14.6 Feed Water Quality,” page 75).
t NOTICE Water leakage can occur if feed water tubing is not watertight! If the
feed water tubing is deformed or not inserted deep enough: Water can leak out.
After startup, make sure that all external water connections are leak proof.

Table of Contents