42 Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential
Startup
6.8.2 Performing a Rinse Cycle
During rinsing, the device fills and rinses the ultrapure water cartridge.
Air is removed from the ultrapure water circulation in the process.
Procedure
t Press the dispense tube into the quick connector of the water outlet.
t Direct the free end of the dispense tube to the drain or into a vessel
(at least 6 liters).
t Start the rinse cycle for the ultrapure water cartridge.
To do this, tap the [START] button.
y During the rinse cycle, the remaining rinse time is displayed in minutes.
t To interrupt the rinse cycle:
t Tap the [x] button.
y The dialog box “Start rinsing” appears again.
t To resume the rinse cycle after an interruption: Tap the [START] button.
y The rinse cycle is performed.
6.8.3 Re-filling the Bag During the Rinse Cycle (only arium
®
mini)
If the bag does not contain enough water or empties during the rinse cycle:
The bag must be refilled.
Procedure
y The dialog box “Fill bag” appears: Fill the bag (see Chapter 6.8, page 41).
1
Start rinsing
Tap “Start” to begin the rinsing process.
w
Rinsing
4:00 min
1
Fill bag
Make sure there is sufcient water to
ll the bag. Tap “START” to begin lling
the bag.
w