44 Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential
Startup
6.9 Connecting the Final Filter
Requirements
The “Connect filter” dialog box appears on the display.
Procedure
t Attach the bell assembly to the final filter.
t Press the final filter into the quick connector of the water outlet.
t Confirm that the final filter has been connected with the [OK] button.
y The display changes to the dispensing screen.
6.10 Rinsing the Final Filter
Requirements
The device is now in dispensing mode.
Procedure
t Place a vessel under the final filter.
t If a sterile filter is used as the final filter: Open the final filter’s vent valve.
t Remove the protective cap from the bell assembly.
t If a sterile filter is used as the final filter: Remove at least 5 liters of water.
This will rinse the final filter.
t If an ultrafilter is used as a final filter: Remove at least 20 liters of water.
This will rinse the final filter.
t Only arium
®
mini plus or arium
®
mini: If the bag does not contain enough water
or empties during the rinse cycle:
t Wait until the bag has filled to a sufficient level.
t Complete the rinse cycle.
t When the rinsing process is complete: Close the vent valve.
t Attach the protective cap to the bell assembly.
y Startup is complete.