46 Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential
Operation
7.2.2 Manually Dispensing Ultrapure Water
During manual dispensing, ultrapure water is dispensed until stopped manually.
Only arium
®
mini plus or arium
®
mini: When the maximum filling level of the bag
has been reached: Dispensing will end automatically.
Procedure
t Tap the [&] symbol.
y Dispensing begins. The volume dispensed thus far is displayed in increments of
0.05 liters (50 ml).
y The ultrapure water flows into the vessel at a maximum throughput rate of
about 1.0 l/min.
y If an arium
®
mini plus or arium
®
mini is used: The filling level of the bag will be
updated on the display during dispensing.
t To stop dispensing: Tap the [STOP] button.
t When dispensing is finished: Attach the protective cap to the final filter bell
assembly.
7.2.3 Volume-controlled Dispensing of Ultrapure Water
During volume-controlled water dispensing, a previously specified amount of water
is dispensed. The volume to be dispensed must be entered with the following
specifications:
The minimum dispensing volume is 0.05 liters (50 ml).
The maximum dispensing volume is 5.00 liters.
Only volumes between 0.05 and 5.00 liters can be entered. Invalid numeric fields
become inactive (white background).
Only arium
®
mini plus or arium
®
mini: When a higher dispensing volume is
entered than that in the bag, dispensing cannot be started. An appropriate
message will be displayed.
Procedure
t Tap the [
$] symbol.
y The numeric keypad for inputting volumes appears.
$&
0.25
0.055 μS/cm c
21.3 °C
5l
3l
1l
0.055 μS/cm c
21.3 °C
STOP
5l
3l
1l
0.90l
$&
0.25
0.055 μS/cm c
21.3 °C
5l
3l
1l