Cleaning and Maintenance
8.2 Maintenance Schedule
Depending on the volume of water dispensed, it may be necessary to change the
consumables more often than specified in the maintenance schedule. If, for example,
sterile water is required, the final filter must be replaced regularly.
It is advisable to replace different consumables in the same maintenance step. This
saves time and water. We recommend that you make a sensible plan for replacing
consumables.
Interval Component Task Chapter, page
12 months UV lamp (optional) Changing the UV Bulb 8.5.3, page 58
Max. 6 months (depending on the volume
of water dispensed)
Ultrapure water cartridge
(Scientific Pack)
Replace ultrapure water cartridge 8.5.4, page 59
6 months
(only arium
®
mini plus or arium
®
mini)
Bag Replacing the Bag 8.5.2, page 56
6 months (only arium
®
mini plus) Pretreatment cartridge Replace the pretreatment cartridge 8.5.3, page 58
1 – 24 weeks (depending on the application)
Sterile filter (final filter) Change final filter 8.5.6, page 62
1 – 13 weeks (depending on the application) Ultrafilter (final filter) Change final filter 8.5.6, page 62
8.3 Opening the “Care” Menu
The “Care” menu contains all the steps for the care and maintenance work.
Procedure
t Tap the
[Care] symbol in the menu.
y The “Care” menu appears. The following tasks can be carried out.
Possible device tasks Chapter, page
Display reminders 8.4, page 55
Replace consumables 8.5, page 55
Enable, disable or configure reminders for
replacing final filters
8.6, page 63
Manual depressurization 8.7, page 63
Carry out venting 8.8, page 64
Dispense Care
Fill bag
Care
Reminder
Change material
Final lter reminder
Depressurization
Ventilation