56 Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential
Cleaning and Maintenance
8.5.2 Replacing Bag (only arium
®
mini plus or arium
®
mini)
Draining the Bag
Requirements
− In the “Change material” menu, the menu item [
Bag] is activated.
− The “Empty bag” dialog box appears.
Procedure
t Check whether the tank outlet tubing is connected to the “Bag Outlet” connection.
t If required: Connect the tank outlet tubing.
t Guide the tank outlet tubing into a suitable container.
t In order to empty the bag completely: Place the container underneath the device.
t Open the ball cock of the tank outlet tubing.
y Pure water flows out of the bag.
t When the bag is empty: Confirm the emptying of the bag with the [
OK] button.
y The dialog box “Change bag” appears.
Removing Empty Bag
Procedure
t Remove the left side cover:
t Reach behind the side cover and pull the side cover to the side (1).
t Pull the side cover up and off to remove it (2).
t To make it easier to reach the lower connections on the device:
Pull the tray up and remove it from the device.
t Disconnect the gray quick connector on the three device connections
successively.
t Slide the empty bag out of the guide rail and remove it from the device.
1
Drain the bag
Open the tank outlet (Bag Outlet) at the
rear of the device.
OK
1
Bag replacement
Remove the left cover of device. Insert
the empty bag as described in the
manual. Put the left cover back on.
OK
2
1