Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential 61
Cleaning and Maintenance
t Disconnect the black connector from the old UV lamp.
t Unscrew the black plastic cover of the UV lamp housing. If required:
Use a suitable tool to help you do this.
t Remove the black plastic cover of the UV lamp housing.
t Carefully pull the old UV lamp out of the lamp housing, package it and safely
dispose of it (see Chapter “13 Disposal,” page 71).
Inserting the UV Lamp
NOTICE
Touching UV lamps with your bare fingers will cause them to become defective!
Touching the UV lamp with your bare fingers will leave fingerprints. The fingerprints
can become so hot during operation that the UV lamp is destroyed.
t Never touch the glass of the UV lamp with your bare fingers.
t Only hold the UV lamp where it connects to the device or when wearing gloves.
Procedure
t Carefully unpack the new UV lamp without touching the glass with your fingers.
t Insert the new UV lamp all the way into the lamp housing without exerting
pressure on it.
t Unscrew the black plastic cover of the UV lamp housing by hand.
t Attach the black connector to the UV lamp. The connector only fits in two
orientations. Both orientations allow the UV lamp to function.
t Slide any protruding cable back into the housing.
t Slide the metal retaining clip over the cable and onto the black plastic cover of
the UV lamp housing.
t Insert the closed side of the metal retaining clip into the recess of the black
plastic cover provided.
t Squeeze the open side of the metal retaining clip and insert it into the recesses
provided in the black plastic cover.
y Release the metal retaining clip secures it to the black plastic cover.
t Connect the device to the power supply.
y The wizard continues.
t Follow the instructions on the display, e.g. “Insert the new pretreatment cartridge
(R) as described in the manual”.