66 Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential
Malfunctions
9 Malfunctions
9.1 Error Messages
If an error message is active, dispensing is canceled and locked automatically.
Error message Fault Cause Correction Chapter, page
Error 0105
Measured values
are not being
displayed.
There is an internal
communications
error.
Disconnect the device from the AC power
and wait a minute.
Reconnect the device to the AC power.
7.1, page 45
If the error persists: Contact Sartorius Service.
Error 0140
Check UV lamp,
replacement may be
needed.
The UV lamp is
not recognized.
The UV lamp is not
connected properly
or defective.
Check whether the black plug of the
UV lamp is connected correctly.
8.5.5, page 60
Remove the UV lamp and check for damage.
If required: Change the UV lamp.
8.5.5, page 60
If the error persists: Contact Sartorius Service.
Error 0150
Check device for
leakage.
Please contact your
service technician.
The device has
a leak.
Water has leaked
inside the device.
Remove the left side cover and check that
the three bag hoses are properly connected
to the device connectors. If required:
Remove the bag and reconnect it.
8.5.2, page 56
Drain the tray at the bottom of the device
and check that the connections are tight
during operation.
If the error persists: Contact Sartorius Service.
Error 0160
Front cover is open.
Please put the front
cover back on.
Operation of the
device is not
possible.
The front cover is
not properly
attached.
Correctly attach the front cover to the device.
If the error persists: Contact Sartorius Service.
Error 0163
Check proper fit of the
pretreatment cartridge
(R).
The pretreatment
cartridge is not
recognized.
The pretreatment
cartridge has not
been properly
inserted.
Press the pretreatment cartridge in firmly
until you hear a distinct clicking sound.
If the error persists:
− Remove the pretreatment cartridge.
− Insert the pretreatment cartridge.
If the error persists: Contact Sartorius Service.
8.5.3, page 58
Error 0166
Check proper fit of the
ultrapure water
cartridge (L).
The ultrapure
water cartridge
is not detected.
The ultrapure water
cartridge has not
been properly
inserted.
Press the ultrapure water cartridge in firmly
until you hear a distinct clicking sound.
If the error persists:
− Remove the ultrapure water cartridge.
− Insert the ultrapure water cartridge.
If the error persists: Contact Sartorius Service.
8.5.4, page 59