Malfunctions
Error message Fault Cause Correction Chapter, page
Error 0180
Please contact your
service technician.
The fill level of
the bag is not
being properly
detected.
The bag is damaged
or improperly
connected.
Remove the left side cover and check the
bag for damage. If required: Change bag.
8.5.2, page 56
Check that the three bag hoses are properly
connected to the device connectors.
If required: Remove the bag and reconnect it.
8.5.2, page 56
The sensor is
defective.
Using the tank outlet tubing connected to
“Bag Outlet”, let about 4 liters of pure water
out of the bag and check the level change.
7.2.6, page 48
If the error persists: Contact Sartorius Service.
Error
The conductivity of
the ultrapure water
is outside of the
measuring range.
The conductivity
of the ultrapure
water is outside
of the measuring
range.
There is air in the
ultrapure water
circulation.
Initiate venting. 8.8, page 64
The ultrapure water
cartridge has been
used up.
Replace the ultrapure water cartridge. 8.5.4, page 59
To dispense water for testing purposes:
Disable the dispensing lock.
If the error persists: Contact Sartorius Service.
Error
The temperature of
the ultrapure water is
outside of measuring
range.
The temperature
of the ultrapure
water is outside
of measuring
range.
The temperature of
the ultrapure water
is outside of
measuring range.
Check whether the ambient temperature
complies with the device specifications in
the technical data.
14.2, page 73
Dispense about 1 liter:
To dispense water for testing
purposes: Disable the dispensing lock.
Dispense and discard about 1 liter of
water.
If the error persists: Contact Sartorius Service.
Error
Ultrapure water quality
> XX µS/cm
The limit has
been exceeded
and the
dispensing lock is
preventing
further
dispensing.
There is air in the
ultrapure water
circulation.
Perform venting. 8.9, page 63
Replace the ultrapure water cartridge. 8.6.3, page 63
To dispense water for testing purposes:
Disable the dispensing lock.
If the error persists: Contact Sartorius Service.
The limit has been
configured
incorrectly.
Check the limit. If required:
Configure the limit.