EasyManua.ls Logo

Sartorius H2O-MA-UV-T - Additional Faults

Sartorius H2O-MA-UV-T
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions arium
®
mini plus | arium
®
mini | arium
®
mini essential 69
Malfunctions
Warning message Fault Cause Correction Chapter, page
Replacement of the
pretreatment cartridge
(R) is required.
The pretreatment
cartridge needs
to be replaced.
The replacement
interval of the
pretreatment cartridge
has expired.
Replace the pretreatment cartridge. 8.5.3, page 58
Replacement of the
ultrapure water
cartridge (L) is required.
The ultrapure
water cartridge
needs to be
replaced.
The replacement
interval of the ultrapure
water cartridge has
expired.
Replace the ultrapure water cartridge. 8.5.4, page 59
Replacement of the bag
is required.
The bag needs
to be replaced.
The replacement
interval of the bag has
expired.
Replace the bag. 8.5.2, page 56
Replacement of the
UV lamp is required.
The UV lamp
needs to be
replaced.
The replacement
interval of the UV lamp
has expired.
Change UV lamp. 8.5.5, page 60
Replacement of the
sterile final filter is
required.
The sterile final
filter needs to be
replaced.
The replacement
interval of the sterile
final filter has expired.
Replace the final filter. 8.5.3, page 58
Sartorius maintenance
is required.
Maintenance
service must be
performed.
The maintenance
service interval has
expired.
Contact Sartorius Service.
9.3 Additional Faults
Fault Cause Correction Chapter, page
The device unexpect-
edly stops dispensing.
The bag is empty. Check the level of the bag on the display.
Remove the left cover and check the fill
level of the bag. If required: Manually fill
the bag (only arium
®
mini).
7.3, page 48
Check that the feed water tubing has been
correctly connected.
6.6.1, page 38
There is no feed water is connected. Connect the feed water tubing or tank
filling tubing.
6.6, page 38
The final filter is clogged or contains air. If the final filter is attached and no water
can be dispensed:
Rinse the final filter.
Vent the final filter.
If the error persists: Replace the final
filter.
6.10, page 44
8.8, page 64
8.5.3, page 58
If the final filter has been removed and no
water can be dispensed: Contact Sartorius
Service.

Table of Contents