Technical Specifications
14.5.2 arium
®
mini and arium
®
mini essential
Unit Value
Water type Ultrapure water ASTM Type 1
Water dispensing flow rate
2
l/min
< 1.0
Volume-controlled dispensing
2
l Between 0.05 and 5 (in 50-ml increments)
Typical conductivity μS/cm 0.055 (compensated to 25 °C
5
)
Typical resistance
MO × cm
18.2 (compensated to 25 °C
5
)
TOC content
3
(system with UV lamp) ppb < 5
Microorganism content
4
CFU/1.000 ml < 1
Particle content > 0.2 μm
4
ml
-1
< 1
2
Depending on the hydrostatic pressure, and connected accessories and/or final filter
3
Determined with municipal water (Goettingen), TOC approx. 1000 ppb
4
When using an arium
®
SterilePlus final filter (Sartopore
®
2 150)
5
Measurement output adjustable compensated to 25 °C or uncompensated
14.6 Feed Water Quality
14.6.1 arium
®
mini plus
Unit Value
Suitability / type
Exclusively potable tap water pursuant to the drinking water
standards of the USA, the European Union or Japan.
Inlet pressure bar 0.5 to 6 (recommended: > 2)
Temperature °C 2 to 30
TOC ppb < 2,000
Max. total hardness (max. CaCO
3
) ppm 360
Free chlorine ppm < 4
Iron (total Fe content) ppm < 0.1
Fouling Index (SDI) < 10
Turbidity NTU < 1
pH value 4 to 10