Setting the Drying Parameters
Select PRG in the function line to adapt parameters for the drying program to the particular
requirements of the product sampled.
Drying Parameters
Temperature during heating
40 … 160°C
End of analysis
0.0 min Select 0.0 minutes for fully automatic shutoff
0.1 to 99 min Select an interval from 0.1 to 99 minutes to define a specific
analysis time
Display mode for result
%M Moisture
%S Dry weight
%MS Ratio
g Residual weight
Start of analysis
E With stability, after the
key is pressed
A Without stability, after the cover is closed
Print intermediate results
0.0 min Off
0.1 to 10.0 min