30
External Calibration and Adjustment with a Factory-Defined Weight
Externally calibrate and adjust the weighing system using a 30-g calibration weight
Step Key (or instruction) Display
1. Turn on the analyzer e Self-test runs
2. Select CAL for calibration/
adjustment
3. Confirm CAL
4. Confirm again when Pb
is shown
5. Tare the weighing system
6. Select CAL again
7. Confirm CAL