EasyManua.ls Logo

Sartorius ME215S - Data Output Functions

Sartorius ME215S
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
123
Data Output Functions
Line for metrological data
Bar graph
Measured value line
Text line
Soft key lables
Plus/minus sign Unit
Stability indicator Tare memory
Calculated value
Application pictograms
Max 210 g
Mi n 1 g
e= 0.1mg
d=0.01mg
J
F H
S
There are 3 options for data output:
Output to the display and control unit
Output to a printer (generate a printout)
Output to a peripheral device (e.g., computer) via the interface port
Output to the Display and Control Unit
The display is divided into 9 sections. Information about the balance,
the application being used and the sample weighed is output in the
following sections:
Line for metrological data
Bar graph
Plus/minus sign, stability symbol display
Line for measured values (weights)
Weight unit display
Data in tare memory; calculated value
Application symbol display
Text line
Soft key labels
Line for Metrological Data (on balances verified for legal metrology)
This line shows:
Maximum weighing capacity (such as 210 g)
Minimum weighing capacity; the weight must not go below this limit
in Germany (such as 1 g)
Verification scale interval; irrelevant if the balance is not used in legal metrology
(such as 0.1 mg)
Readability/actual scale interval: indicates the balance’s display increment in digits
(such as 0.01 mg)
Bar Graph (overview display)
In the bar graph, weighing results are displayed either
as a percentage of the maximum balance capacity, or
in relation to a target value, with tolerance limits indicated.
You can turn off (blank) the bar graph display (Setup: Device parameters:
Display: Digit size: 13 mm + text display or 13 mm)
Plus/Minus Sign, Stability Symbol
This section shows:
“Busy” symbol
Plus or minus sign
Zero symbol (indicating the scale has been zeroed)

Table of Contents