13
Calibration
Unload the balance. When the display reads zero, press the
tare control
(10)
for at least three seconds until the calibration
weight appears in the weight display.
Center the calibration weight (see Accessory Kit) on the pan
and dose the weighing chamber.
The weight unit symbol
"mg"
will now appear in the weight dis-
play. An acoustic Signal indicates the end of the calibration pro-
cedure.
You can block the calibration function – see "Balance Operating
Program." This function is accessible whenever the balance op-
erating program is unlocked by the menu access switch.
Make sure to calibrate the balance each time you exchange
the Standard pan for the one for weighing filaments and
vice versa to rule out any effects caused by the inevitable
differences in weights of the pans.