21
Specifications
Model M 500 P
Capacity / weighing range mg 52/102/202/510
Readability mg 0,002/0,005/0,01/0,02
Tare range (by subtraction) mg 510
Standard deviation mg
≤±0,004/0,005/0,01/0,02
Max. linearity mg
≤±0,004/0,01/0,02/0,04
Stabilization time (typical) s 8
Display update s 0,1 – 0,8 (selectable)
Adaption to ambient conditions
and application requirements
By selection of 1 of 4 optimized filter levels
Stability range d 0,25 ... 64 (selectable)
Ambient temperature range
o
C +15 to +30
Sensitivity drift within 290 … 300 K /K
≤± 1∙10
-5
Pan size mm
mm
28 Ø (standard pan)
28 Ø (special pan for weighing filaments)
Balance housing (W x D x H) mm 219 x 291 x 135
Weighing chamber (W x D x H) mm 54 x 53 x 50
Net weight, approx. kg 547
Netzspannung, Frequenz 50-60 Hz
115 oder 230 V entsprechend verwendetem Netzteil
Leistungsaufnahme VA 7,2 (typisch)
Schnittstelle
RS 232 C/V24 – 28, RS 423/V10; 7-bit;
parity: even, mark, odd, space;
Übertragungsgeschwindigkeit 150 ... 9600 Baud