Error Messages
Display Cause Solution
H Scale capacity exceeded Unload the scale
L No pan on the scale Place the pan on the scale
Something is touching the Remove whatever is touching
pan the pan
E 01 Display overflow; i.e., the value to Reduce the load on the scale
be output cannot be displayed
E 02 Calibration parameter not met; Only start calibration when zero
e.g.: is displayed
– scale not zeroed Press = to zero the scale
– load on scale Unload the scale
E 08 Scale is not within the zero range Scale can only be zeroed setting
within the zero-setting range
(± 2% of max. load)
E 09 Scale cannot be tared when the Press = to zero the scale
gross value is ≤ 0
E 10 = key is blocked when there is Press c to clear the 2nd tare
data stored in the 2nd tare memory memory
E 11 Value cannot be stored in the Press =
2nd tare memory
E 22 Weight too light, or no load Increase the weight on the scale
on the scale
E 30 p key pressed while BPI mode Set the scale to the SBI mode
active (e.g., by activating the “Close”
function in the MC1 server)
E 50 Overflow or underflow in the TK sensor or main PCB is defective
temperature compensation converter (replacement required)
E 53 Temperature compensation converter TK sensor or main PCB is defective
is not functioning. The processor (replacement required)
is not receiving any measured data
from the TK switch
E 54 A/D converter output is below The scale is underloaded; either
the minimum limit the pan is not on the scale
or there is a mechanical defect
or a defect in the main PCB
E 55 A/D converter output is above the Either the scale is overloaded
maximum limit or there is a mechanical defect
or a defect in the main PCB