Profibus Interface PR 5210 Operating Manual
62 Sartorius
6.3.3.4 Parameter P24: SetSpan
Write only
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 Weight MSB
Byte 1 ""
Byte 2 Weight LSB
Byte 3 Exponent Unit
Calibrating with weights
Weight 3 Bytes for the weight value
Exponent Number of digits behind the decimal comma/point
Unit Weight unit (as in register 4)
6.3.3.5 Parameter P25: Set/GetDeadloadMvpv
Write and read
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 DEADLOAD MSB
Byte 1 ""
Byte 2 ""
Byte 3 DEADLOAD LSB
DEADLOAD in mV/V (value in mV/V )*10
6
6.3.3.6 Parameter P26: Set/GetSpanMvpv
Write and read
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 SPAN MSB
Byte 1 ""
Byte 2 ""
Byte 3 SPAN LSB
SPAN in mV/V (value in mV/V )*10
6
6.3.3.7 Parameter P27: CalcTest
Write only
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 1
Calculate testvalue