PR 5210 Operating Manual Setup and Configuration
Sartorius 27
4.8.1.2 Analog range
The following selections are possible:
0-20 mA
Analog output range
4-20 mA
Analog output range
4.8.1.3 Analog error
The following selections are possible:
hold
In case of error the analog output keeps the last value
0 mA
In case of error the analog output goes to 0 mA
4 mA
In case of error the analog output goes to 4 mA
20 mA
In case of error the analog output goes to 20 mA
4.8.1.4 Analog < 0
Behaviour when the weight value is blow zero. The following selections are possible:
linear
In case of negative weight the analog output continues (only possible if the output
value for zero weight is larger than 0 mA)
0 mA
In case of negative weight the analog output goes to 0 mA
4 mA
In case of negative weight the analog output goes to 4 mA
20 mA
In case of negative weight the analog output goes to 20 mA
4.8.1.5 Analog > FSD
Behaviour when the weight value is above FSD. The following selections are possible:
linear
In case of weight above end of the scale the analog output continues (only possible
if the output value for FSD is smaller than 20 mA)
0 mA
In case of weight above end of the scale the analog output goes to 0 mA
4 mA
In case of weight above end of the scale the analog output goes to 4 mA
20 mA
In case of weight above end of the scale the analog output goes to 20 mA
4.8.1.6 Analog value
A fixed analog value can be entered to set the analog output. The analog mode has to be set to:
transparent. If the PLC has written the analog value it is displayed in the configuration tool.
4.8.1.7 Weight for 0/4 mA
Enter the weight value at which the analog output shall show 0 mA (or 4 mA, if analog range is set to
4 –20 mA)
4.8.1.8 Weight for 20 mA
Enter the weight value at which the analog output shall show 20 mA.
4.8.2 Profibus Address
The address on the profibus has to be defined here, valid addresses are in the range 1, 2 ... 126.
Default is 10.