PR 5210 Operating Manual Setup and Configuration
Sartorius 31
4.10 Analog Output Adaption
For PR 5210/00 und PR 5210/10 only.
The analog output current is at the receiver side usually routed over a resistor, measured as voltage
and the digitalisied. The errors of this transfer chain could be compensated at the transmitter side.
With this menu the points for 4 and 20 mA could be individually adapted, so that they are precise
again at the receiver side.
The adaption is done in several steps:
1. step: Press [Start Adaption].
2. step: The analog output is set to 4mA. Measure the current at the receiver side and enter the
value of the measured current. E.g. 4.002mA.. Press [next].
3. step: The analog output is set to 20mA. Measure the current at the receiver side and enter the
value of the measured current. E.g. 20.003mA.. Press [next].
The current is now corrected to 4.000mA and 20.000mA.