EasyManua.ls Logo

Sartorius PR 5210/10 - Scale Information I and N Command Response

Sartorius PR 5210/10
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PR 5210 Operating Manual SMA Protocol
Sartorius 39
“MOD” product model identification
(response of 2nd ‘B’ command)
“REV” software revision
(response of 3rd ‘B’ command)
“SN_“ serial number
(response of 4th ‘B’ command)
“END” this is always the last About field
(response of the last ‘B’ command)
‘:’ Separator between field name and field contents.
<yyyyyy> Data field contain 25 characters maximum.
SMA field contents <level/revision>
where: level= (1, 2, etc.); revision= (1.0, 1.1, etc.)
<CR> End of About response
Example:
Command Response
<LF> A <CR> <LF>SMA:1/1.0 <CR>
<LF> B <CR> <LF>MFG:Sartorius <CR>
<LF> B <CR> <LF>MOD:PR5210 <CR>
<LF> B <CR> <LF>REV:02.09.9 <CR>
<LF> B <CR> <LF>SN_:148388723 <CR>
<LF> B <CR> <LF>END: <CR>
Note: if the host should ask for additional ‘B’ status then.
<LF> B <CR> <LF> ? <CR>
5.4.6 Scale Information ‘I’ and ‘N’ Command Response
Response Format for Information commands ‘I’, ‘N’ (variable length):
<LF><xxx>:<yyyyyy><CR>
where:
<LF> Start of Information response
<xxx> About field descriptor is fixed at 3 characters, is left justified, filled with blanks on the
right side.
Following fields are required:
“SMA” compliance level/revision
(response of ‘I’ command)
“TYP” Scale type: ‘S’= Scale
(response 1st ‘N” command)
“CAP” capacity of range, unit-of-measure, count-by and decimal position,
each delimited by ‘:’
where: yyyyyy= uuu:c..c:n:d
uuu= unit-of-measure
c..c= full scale capacity of this range
(may include decimal point)
n= least significant count-by digit
(e.g. 1, 2, 5,10,20…) no decimal point
d= decimal point position
‘0’= none
‘1’= xxxx.x
‘2’= xxx.xx
‘3’= xx.xxx . . etc.

Table of Contents

Related product manuals